首页 | 本学科首页   官方微博 | 高级检索  
     


3D Cube FLAIR plus HyperSense compressed sensing is superior to 2D T2WI FLAIR scanning regarding image quality,spatial resolution,detection rate for cortical microinfarcts
Authors:Tiefang Liu  Yonghao Wang  Zhengyang Xu  Tao Wu  Xiao Zang  Meng Li  Jinfeng Li
Affiliation:a Department of Radiology, The First Medical Center of PLA General Hospital, Beijing, China;b Department of Ultrasound, The Eighth Medical Center of PLA General Hospital, Beijing, China;c GE Healthcare MR Enhanced Application Team, Beijing, China.
Abstract:3-dimention (3D) Cube isotropic volumetric magnetic resonance imaging (MRI) facilitates comprehensive recognition of microinfarcts while it takes long scanning time. HyperSense compressed sensing is an emerging technique for accelerating MRI acquisition to reduce scanning time, while its application along with 3D Cube MRI for microinfarcts is seldom reported. Therefore, this study aimed to investigate the efficiency of 3D Cube FLAIR plus HyperSense compressed sensing technique versus conventional 2-dimention (2D) FLAIR scanning in the detection of cortical microinfarcts (CMIs).Totally 59 patients with cerebrovascular disease were enrolled then scanned by 3D Cube FLAIR plus HyperSense compressed sensing and 2D T2WI FLAIR sequences. The image quality scores, signal-to-noise ratio (SNR) for gray matter (GM), SNR for white matter (WM), their contrast-to-noise ratio (WM-to-GM CNR), detected number of CMIs were evaluated.3D Cube FLAIR plus HyperSense showed a dramatically increased scores of uniformity, artifact, degree of lesion displacement, and overall image quality compared to 2D T2WI FLAIR. Meanwhile, it exhibited similar SNRwm and SNRgm, but a higher WM-to-GM contrast-to-noise ratio compared with 2D T2WI FLAIR. Furthermore, the scanning time of 3D Cube FLAIR plus HyperSense and 2D T2WI FLAIR were both set as 2.5 minutes. Encouragingly, 244 CMIs were detected by 3D Cube FLAIR plus HyperSense, which was higher compared to 2D T2WI FLAIR (106 detected CMIs).3D Cube FLAIR plus HyperSense compressed sensing is superior to 2D T2WI FLAIR scanning regarding image quality, spatial resolution, detection rate for CMIs; meanwhile, it does not increase the scanning time. These findings may contribute to early detection and treatment of stroke.
Keywords:2D T2WI FLAIR   3D Cube FLAIR   cortical microinfarcts   HyperSense compressed sensing   magnetic resonance imaging
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号