首页 | 本学科首页   官方微博 | 高级检索  
检索        


Long-term biocompatibility of implanted polymer-based intrafascicular electrodes
Authors:Lawrence Stephen M  Larsen Jytte O  Horch Kenneth W  Riso Ron  Sinkjaer Thomas
Institution:Department of Bioengineering, University of Utah, Salt Lake City, Utah 84112, USA.
Abstract:Polymer-based longitudinal intrafascicular electrodes (polyLIFEs) were chronically implanted into the sciatic nerve of white New Zealand rabbits (n=8) for a period of 6 months (hereafter referred to as the long-term group). The impact of the implantation procedure, as observed 6 months post surgery, was evaluated in a sham-treated control group (n=9). The contralateral sciatic nerve served as the control for each animal. Nerve-fiber counts, fiber diameters, and myelin thickness were estimated at the level of the implant site, 1.5 cm proximally, and 1.5 cm distally for both nerves in sham-treated and long-term groups. Implantation of polyLIFEs had no significant effect on fiber counts, nerve-fiber diameter, or myelin thickness. A slight increase in connective tissue in the vicinity of the implant site was evident in the long-term group, including a thin but dense capsule immediately surrounding the implanted electrode.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号