首页 | 本学科首页   官方微博 | 高级检索  
检索        


Frequency measurements in real-time US equipment: variations from expected values
Authors:Kimme-Smith  C; Duerinckx  AJ; Kaiser  S; Jones  JP; Tessler  FN; Bassett  LW
Institution:Department of Radiological Sciences, UCLA School of Medicine 90024.
Abstract:The frequency responses of nine real-time mechanical probes and two pulsed Doppler probes from four manufacturers were measured. The reflected frequency was measured in a clinical environment by evaluating the pulser, transducer assembly, and receiver as a system. Two independent systems were used for these measurements: a spectrum analyzer and a data acquisition system that computed the fast Fourier transform of the reflected waveform. Results showed that probe frequency (calibrated by the manufacturer in the transmit mode with a hydrophone) was as much as 30% higher than that found by means of frequency measurement procedures with the probe attached to the ultrasound system and evaluated in the receive mode. Such large discrepancies indicate that if acceptance testing of the frequency spectrum is not performed at a clinical facility, prolonged clinical evaluation of new equipment should be arranged.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号