Dosimetry in MARS spectral CT: TOPAS Monte Carlo simulations and ion chamber measurements
Authors:Gray Lu  Steven Marsh  Jerome Damet  Pierre Carbonez  John Laban  Christopher Bateman  Anthony Butler  Phil Butler
Institution:1.University of Canterbury,Christchurch,New Zealand;2.CERN,Geneva,Switzerland;3.Institute of Radiation Physics,Lausanne University Hospital,Lausanne,Switzerland;4.Department of Radiology,University of Otago,Christchurch,New Zealand;5.National Centre for Radiation Science,Christchurch,New Zealand;6.MARS Bioimaging Ltd,Christchurch,New Zealand
Abstract:Spectral computed tomography (CT) is an up and coming imaging modality which shows great promise in revealing unique diagnostic information. Because this imaging modality is based on X-ray CT, it is of utmost importance to study the radiation dose aspects of its use. This study reports on the implementation and evaluation of a Monte Carlo simulation tool using TOPAS for estimating dose in a pre-clinical spectral CT scanner known as the MARS scanner. Simulated estimates were compared with measurements from an ionization chamber. For a typical MARS scan, TOPAS estimated for a 30 mm diameter cylindrical phantom a CT dose index (CTDI) of 29.7 mGy; CTDI was measured by ion chamber to within 3% of TOPAS estimates. Although further development is required, our investigation of TOPAS for estimating MARS scan dosimetry has shown its potential for further study of spectral scanning protocols and dose to scanned objects.