首页 | 本学科首页   官方微博 | 高级检索  
     


New FOCI pulses with reduced radiofrequency power requirements
Authors:Shen Jun  Chen Zhengguang  Yang Jehoon
Affiliation:Molecular Imaging Branch, National Institute of Mental Health, Bethesda, Maryland, USA. shenj@intra.nimh.nih.gov
Abstract:PURPOSE: To propose new frequency offset corrected inversion (FOCI) pulses with significantly reduced radiofrequency (RF) power deposition for spin echo imaging by incorporating the variable-rate selective excitation (VERSE) schemes into the pulse design. MATERIALS AND METHODS: Two schemes are proposed to design the new FOCI pulses with dramatically reduced peak RF power requirements. In scheme A, the time-dilation function is derived from a predefined adiabaticity factor modulation function. In scheme B, the time-dilation function is predefined, while the adiabaticity factor is conserved. RESULTS: The new FOCI pulses are shown to be able to operate at reduced specific absorption rate (SAR), specifically at the same peak RF power as that of a five- or seven-lobe sinc inversion pulse of the same duration. Using the new FOCI pulse, significant gain in sensitivity was observed in in vivo spin-echo echo-planar imaging, which was attributed to the improved refocusing slice profile. CONCLUSION: The new FOCI pulses can replace the 180 degrees five- or seven-lobe sinc pulses in spin-echo imaging with the same peak RF power requirement and significantly improved slice profile.
Keywords:RF pulse  RF power deposition  adiabaticity  spin echo imaging