A B(0) shift correction method based on edge RMS reduction for EPI fMRI
Authors:Kochunov P V  Liu H L  Andrews T  Gao J H  Fox P T  Lancaster J L
Institution:Research Imaging Center, The University of Texas Health Science Center, San Antonio, Texas 78284, USA.
Abstract:Shifting of echoplanar images (EPI) in the phase-encoding direction during functional magnetic resonance imaging (fMRI) experiments can be observed due to B(0) drift. These shifts can cause artifacts in functional activation maps that can be corrected using a navigator echo (NE) technique, but the NE correction requires pulse sequence modifications not available on many clinical systems. A fast, postprocessing correction method based on edge root-mean-square error reduction (ERMSR) is introduced and shown to provide an equivalent correction. J. Magn. Reson. Imaging 2000;12:956-959.