Comparison of Two Types of Neonatal Extracorporeal Life Support Systems With Pulsatile and Nonpulsatile Flow
Authors:Nikkole Haines  Shigang Wang  John L Myers  Akif Ündar
Institution:1. Pediatric Cardiac Research Laboratories, Department of Pediatrics, Penn State Milton S. Hershey Medical Center, Penn State Hershey College of Medicine, Penn State Hershey Children's Hospital;2. and;3. Department of Surgery and Bioengineering, Penn State Milton S. Hershey Medical Center, Penn State Hershey College of Medicine, Penn State Hershey Children's Hospital, Hershey, PA, USA
Abstract:We compared the effects of two neonatal extracorporeal life support (ECLS) systems on circuit pressures and surplus hemodynamic energy levels in a simulated ECLS model. The clinical set‐up included the Jostra HL‐20 heart–lung machine, either the Medtronic ECMO (0800) or the MEDOS 800LT systems with company‐provided circuit components, a 10 Fr arterial cannula, and a pseudo‐patient. We tested the system in nonpulsatile and pulsatile flow modes at two flow rates using a 40/60 glycerin/water blood analog, for a total of 48 trials, with n = 6 for each set‐up. The pressure drops over the Medtronic ECLS were significantly higher than those over the MEDOS system regardless of the flow rate or perfusion mode (144.8 ± 0.2 mm Hg vs. 35.7 ± 0.2 mm Hg, respectively, at 500 mL/min in nonpulsatile mode, P < 0.001). The preoxygenator mean arterial pressures were significantly increased and the precannula hemodynamic energy values were decreased with the Medtronic ECLS circuit. These results suggest that the MEDOS ECLS circuit better transmits hemodynamic energy to the patient, keeps mean circuit pressures lower, and has lower pressure drops than the Medtronic Circuit.
Keywords:Extracorporeal life support systems  Extracorporeal membrane oxygenation  Energy equivalent pressure  Surplus hemodynamic energy (extra energy)
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号