Assessment of stenosis in vascular access grafts |
| |
Authors: | Van Tricht Ilse De Wachter Dirk Vanhercke Daniel Tordoir Jan Verdonck Pascal |
| |
Affiliation: | Hydraulics Laboratory, Institute Biomedical Technology, Ghent University, Ghent, Belgium. Vantricht@ugent.be |
| |
Abstract: | The major complication that occurs with grafts used as vascular access for hemodialysis, is stenosis at the venous anastomosis or in the draining vein. 75% area stenosis is considered significant as thrombotic occlusion may occur. The aim of this experimental study was to evaluate invasive and noninvasive indices to detect significant stenoses in a vascular access graft. A compliant underarm loop graft in vitro model was built and studied with 50, 65, 80, and 90% stenosis at flow rates of 500, 1000, and 1500 mL/min. Flow in the system was pulsatile. Velocity was measured with ultrasound Doppler and the pressure was measured invasively. The resistance index (RI), p(venous line)/MAP, and the newly introduced pressure ratio (PR) were calculated and compared. A stenosis can be suspected when a high frequency ultrasound velocity signal develops at the venous anastomosis. RI > 1 confirms a very severe stenosis (90%). The parameter PR < 8% confirms significant stenoses showing its clinical relevancy. |
| |
Keywords: | In vitro model Hemodialysis Graft Stenosis index |