首页 | 本学科首页   官方微博 | 高级检索  
     


Assessment of stenosis in vascular access grafts
Authors:Van Tricht Ilse  De Wachter Dirk  Vanhercke Daniel  Tordoir Jan  Verdonck Pascal
Affiliation:Hydraulics Laboratory, Institute Biomedical Technology, Ghent University, Ghent, Belgium. Vantricht@ugent.be
Abstract:The major complication that occurs with grafts used as vascular access for hemodialysis, is stenosis at the venous anastomosis or in the draining vein. 75% area stenosis is considered significant as thrombotic occlusion may occur. The aim of this experimental study was to evaluate invasive and noninvasive indices to detect significant stenoses in a vascular access graft. A compliant underarm loop graft in vitro model was built and studied with 50, 65, 80, and 90% stenosis at flow rates of 500, 1000, and 1500 mL/min. Flow in the system was pulsatile. Velocity was measured with ultrasound Doppler and the pressure was measured invasively. The resistance index (RI), p(venous line)/MAP, and the newly introduced pressure ratio (PR) were calculated and compared. A stenosis can be suspected when a high frequency ultrasound velocity signal develops at the venous anastomosis. RI > 1 confirms a very severe stenosis (90%). The parameter PR < 8% confirms significant stenoses showing its clinical relevancy.
Keywords:In vitro model    Hemodialysis    Graft    Stenosis index
本文献已被 PubMed 等数据库收录!