Multicentre evaluation of glycated haemoglobin (HbA1c) of Roche Diagnostics in Andalusia
Authors:Ana Sáez-Benito Godino  José Manuel Vergara Chozas  Apolonia Márquez Ronchel  Fátima Barrero Alor  Lara Papay Ramírez  Nieves María Coronado Álvarez  José de la Torre Fernández  Federico Navajas Luque  Mª Victoria Camacho Reina  Rosario Aguilar Peña  Mª Lourdes de la Peña Carretero  Mª Victoria Romero Sotomayor
Institution:1. H.U. Puerta del Mar, Cádiz, Spain;2. H.U. J.R. Jiménez, Huelva, Spain;3. H.U. San Cecilio, Granada, Spain;4. H. Axarquía, Vélez-Málaga, Spain;5. C.H. Ciudad de Jaén, Jaén, Spain;6. H. Cabra, Córdoba, Spain
Abstract:

Background

A Spanish multicentre evaluation of the third generation of Roche Diagnostics immunoturbidimetric inhibition method (TINIA) is presented for quantification of haemoglobin A1c (HbA1c) in whole blood.

Methods

The analytical performance of the TINIA test was evaluated and blood sample results were compared with two other widely used analysers, Bio-Rad Variant II and Adams Arkray HA-8160, based on HPLC.

Results

Within- and between-batch imprecision (coefficients of variation (CVs)) for HbA1c levels of 5, 6, 7 and 8% were 0.77, 1.23, 1.35 and 1.26% and 2.38, 1.51, 1.76 and 2.16%, respectively. For low (5.4% A1c) and high (10.1% A1c) quality control samples, the within and between-batch %CV were: 1.26; 1.43 and 2; 1.71 respectively. The test met the expected performance in most aspects, except for linearity, that is under the reported range, and HbF interferences, detected for levels over 7.5%. There was a good concordance between the results of TINIA and Variant-IIt in the whole range and with HA-8160 only up to levels of 9%. Between-batch imprecision suggests more frequent calibrations than reported by the provider to maintain variability within the limits established by clinical practice guidelines.

Conclusions

The assay meets the necessary quality standards for routine use, as long as we keep the analytical variability within narrow limits. The results may be interchangeable with the tested HPLC systems, but HbF interference is not detected and it happens at lower levels than reported.
Keywords:Diabetes  Evaluation  Glycated haemoglobin  HbA1c  Immunoturbidimetric inhibition method  Multicentre study