Cooled-Tip Ablation Results in Increased Radiofrequency Power Delivery and Lesion Size in the Canine Heart: Importance of Catheter-Tip Temperature Monitoring for Prevention of Popping and Impedance Rise
Authors:Ichiro Watanabe  Riko Masaki  Nuo Min  Naohiro Oshikawa  Kimie Okubo  Hidezou Sugimura  Toshiaki Kojima  Satoshi Saito  Yukio Ozawa  Katsuo Kanmatsuse
Institution:(1) The Second Department of Medicine, Nihon University School of Medicine, 30-1 Oyaguchi-kami, Itabashi-Ku, Tokyo, Japan, 173-8610
Abstract:Since myocardial lesion size during radio-frequency (RF) ablation is limited at high power by impedance rise when electrode tip temperature exceed 100 °C, controlling tip temperature by continuous intraelectrode saline infusion could permit generation of larger lesion. (1) Two dogs randomly received either standard or cooled tip RF ablation at 4 to 6 separate LV sites. Power output of 30 W was delivered via modified 7 Fr deflectable catheter with 4 mm tip for up to 120 sec or until impedance rise occurred. (2) Six dogs randomly received cooled tip RF ablation at power output of 20, 30, 40 W for 120 sec. (3) Three dogs randomly received cooled tip RF ablation using room temperature saline (21–25 °C) or chilled saline (1–4 °C) infusion. Results: Overall, peak tip temperature was lower for cooled vs standard RF deliveries (97±17 °C vs. 42±8 °C). Lesion depth and volume were significantly larger for cooled burns. Lesion depth and volume and the incidence of abrupt impedance rise/popping did not differ between room temperature saline and chilled saline infusion. The catheter-tip temperature at the onset of popping and abrupt impedance rise was 54±5 °C(48–60 °C) and 59±10 °C(50–75 °C). Conclusion: Cooled tip RF current delivery at high power is associated with increased myocardial lesion size which may facilitate successful ablation of ventricular tachycardia associated with acquired structural heart disease. Catheter-tip temperature should be maintained below 45 °C to prevent popping and abrupt impedance rise during RF energy delivery.
Keywords:catheter ablation  cooled-tip electrode  lesion size  impedance rise  popping
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号