Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines |
| |
Authors: | Mustafa Ata? Süleyman Demircan Arzu Seyhan Karatepe Ha?ha? Ahmet Gülhan G?kmen Zarars?z |
| |
Affiliation: | Department of Ophthalmology and Vision Sciences, Kayseri Traning and Research Hospital, Kocasinan-Kayseri 38010, Turkey;Department of Ophthalmology and Vision Sciences, Kayseri Traning and Research Hospital, Kocasinan-Kayseri 38010, Turkey;Department of Ophthalmology and Vision Sciences, Kayseri Traning and Research Hospital, Kocasinan-Kayseri 38010, Turkey;Department of Ophthalmology and Vision Sciences, Kayseri Traning and Research Hospital, Kocasinan-Kayseri 38010, Turkey;Department of Biostatistics, Faculty of Medicine, Erciyes University, Kocasinan-Kayseri 38010, Turkey |
| |
Abstract: | AIMTo compare and evaluate the phacoemulsification parameters and postoperative endothelial cell changes of two different phacoemulsification machines, each with different modes, but also to assess the relationship between postoperative endothelial cell loss and the phacoemulsification parameters, as well as the other factors in both groups.METHODSThis prospective observational study was comprised of consecutive eligible cataract patients operated with phacoemulsification technique performed by the same surgeon using either a WHITESTAR Signature Ellips FX (transversal, group 1) or Infiniti OZil IP (torsional, group 2) machine.RESULTSThe study included 86 patients. Baseline characteristics in the groups were similar. The median nuclear sclerosis grade was 3 (2-4) in the first group and 2 (2-4) in the second group (P=0.265). Both groups had similar phacoemulsification needle times (group 1: 60.63±36 s; group 2: 55.98±30 s; P=0.789). The percentage of endothelial cell loss 30d after surgery ranged from 3% to 15% with a median of 7% in group 1, and from 2% to 13% with a median of 6% in group 2; however, there was no statistically significant difference between the groups (P=0.407). Hexagonality (P=0.794) and the coefficient of variation (CV; P=0.142) did not differ significantly between the groups before and 30d after surgery. A significant positive correlation was found between the endothelial cell loss and nuclear sclerosis grade (group 1: P<0.001; group 2: P<0.001) and between the endothelial cell loss and average phacoemulsification power (group 1: P=0.007; group 2: P=0.008).CONCLUSIONBoth of these machines were efficient, with similar endothelial cell loss. This endothelial cell loss was related to the increased nuclear sclerosis grade and increased phacoemulsification power. |
| |
Keywords: | endothelial cell changes transversal, torsional phacoemulsification |
|
| 点击此处可从《国际眼科》浏览原始摘要信息 |
|
点击此处可从《国际眼科》下载免费的PDF全文 |
|