Measurements of ultrasonic backscattered spectral centroid shift from spine in vivo: methodology and preliminary results
Authors:Garra Brian S  Locher Melanie  Felker Steven  Wear Keith A
Affiliation:University of Vermont Radiology Department, Burlington, VT, USA.
Abstract:Ultrasonic backscatter measurements from vertebral bodies (L3 and L4) in nine women were performed using a clinical ultrasonic imaging system. Measurements were made through the abdomen. The location of a vertebra was identified from the bright specular reflection from the vertebral anterior surface. Backscattered signals were gated to isolate signal emanating from the cancellous interiors of vertebrae. The spectral centroid shift of the backscattered signal, which has previously been shown to correlate highly with bone mineral density (BMD) in human calcaneus in vitro, was measured. BMD was also measured in the nine subjects' vertebrae using a clinical bone densitometer. The correlation coefficient between centroid shift and BMD was r = -0.61. The slope of the linear fit was -160 kHz / (g/cm(2)). The negative slope was expected because the attenuation coefficient (and therefore magnitude of the centroid downshift) is known from previous studies to increase with BMD. The centroid shift may be a useful parameter for characterizing bone in vivo.
Keywords:Bone   Vertebra   Backscatter