Long-term clinical performance of AAI pacing in patients with sick sinus syndrome: a comparison with dual-chamber pacing.
Authors:Hiroshi Masumoto  Yuichi Ueda  Rinya Kato  Akihiko Usui  Takashi Maseki  Yasushi Takagi  Masato Usui
Institution:Department of Cardio-thoracic Surgery, Nagoya University School of Medicine, 65 Turumai-cho, Syowa-ku, Nagoya City, Japan. hmasu@catvmics.ne.jp
Abstract:AIMS: In this clinical study, we compared two groups of age-matched patients, AAI and DDD, to evaluate the clinical benefits of AAI pacing in patients with sick sinus syndrome (SSS) and normal atrioventricular (AV) conduction. METHODS AND RESULTS: Ninety-five patients with SSS implanted with AAI pacemakers were compared with 101 SSS patients implanted with DDD pacemakers. Mortality, chronic atrial fibrillation, lead survival rates, and reoperation rates were compared by Kaplan-Meier analysis. Eight AAI devices were switched to DDD due to high-degree (grade 2-3) AV block. The incidence of high-degree AV block was 1.104%/year, with a freedom rate of 88.6% at 10 years. There were no significant differences between the two groups in survival rates (87.8% in AAI vs. 93.4% in DDD at 10 years), freedom from atrial fibrillation (93.6% vs. 90.6%), or freedom from reoperation (71.3% vs. 76.3%). On the other hand, lead failure was twice as frequent in the DDD group than in the AAI group (relative risk=2.045, P=0.0382). CONCLUSION: AAI pacing, a simple system using a single lead and single-chamber pacemaker, can achieve a clinical outcome similar to that of the DDD mode in patients with SSS and normal AV conduction.
Keywords:sick sinus syndrome  AAI pacing  DDD pacing  long-term follow up
本文献已被 Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号