Embolization of radiopaque tip component of clip delivery system of MitraClip device: A rare complication with successful percutaneous retrieval
Authors:Ivan P. Casserly MB  BCh  Ernesto Salcedo MD  John Carroll MD
Affiliation:1. Interventional Section, Division of Cardiology, University of Colorado, , Denver, Colorado;2. Imaging Section, Division of Cardiology, University of Colorado, , Denver, Colorado
Abstract:This case report describes a rare complication that occurred during the performance of a MitraClip procedure. Following deployment of the clip, the portion of the device that attaches the clip to the clip delivery system (CDS), referred to as the radiopaque tip, embolized to the left atrium during removal of the CDS through the guide catheter. The percutaneous strategy that was used to successfully retrieve the radiopaque tip is described and discussed.© 2012 Wiley Periodicals, Inc.
Keywords:congestive heart failure  embolization  valvular heart disease