Abstract:The analytical and clinical performances of the new fluorescent immunoassay (CK-MB mass Vidas-BioMerieux) were examined and compared to the chemiluminescent test (CK-MB mass Access-Sanofi-Pasteur). Assay precisions of the CK-MB Vidas test within-assay or between-assay were less than 5.4 and 5.3%, respectively. Linearity was tested up to 214 microg/L. The CK-MB Vidas test was free of interference with CK-BB, CK-MM, and macro-CK. One hundred nineteen blood samples from patients with ischemic myocardial injury (IMI): acute myocardial infarction (AMI), suspected myocardial contusion (SMC), and unstable angina pectoris (UA), were tested using both immunoassays. In AMI, a good correlation was found (Y CK-MB Access] = 1.1372 x CK-MB Vidas] - 6.3902; r(2) = 0.96). In UA and SMC, low values were observed and both methods were well correlated (Y CK-MB Access] = 1.3662 x CK-MB Vidas] + 0.0671; r(2) = 0.97). Clinical data were in good agreement with both immunoassays. ROC analysis performed in AMI demonstrated that the clinical performances of the two assays were similar.