首页 | 本学科首页   官方微博 | 高级检索  
     


Thermosensitive determination of patency in lumboperitoneal shunts. Technical note
Authors:Y Ishiwata  Y Chiba  T Yamashita  G Gondo  K Ide  T Kuwabara
Affiliation:Department of Neurosurgery, Kanagawa Prefectural Ashigarakami Hospital, Japan.
Abstract:Surface cooling and thermistor recording over shunt tubing was used in 23 studies of cerebrospinal fluid shunt patency in 19 patients with lumboperitoneal shunts and normal-pressure hydrocephalus. Shunt patency was shown by downward reflection of the recording trace similar to that obtained for ventriculoperitoneal shunts. Obstruction was demonstrated by a flat-line recording or an upward deflection.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏