Reliability of a tympanic thermometer in measuring temperatures in children after minor ear surgery
Authors:Pandey A  Ingrams D R  Jones M  Raman R  Marks N D
Institution:Department of Otorhinolaryngology, Royal Gwent Hospital, Newport, Wales. anitapandey13@hotmail.com
Abstract:AIM: To evaluate the reliability of infrared tympanic thermometry in children who have undergone myringotomy with grommet insertion. METHOD: Forty children who had undergone myringotomy with at least one grommet insertion had the tympanic temperature of each ear and the axillary temperature measured on admission and 30 minutes post-operatively. RESULT: No difference was found between the pre- and post-operative temperatures measured by either method (p > 0.05, paired t-test, hypothesized difference of 0). CONCLUSION: Infrared tympanic thermometry is reliable in monitoring body temperature in children who have had minor ear surgery.