Errors in low flow oxygen delivery systems
Authors:G. R. WILLIAMS  I. B. MASTERS  M. A. HARRIS
Affiliation:Respiratory Unit, Mater Children's Hospital, South Brisbane, Queensland, Australia
Abstract:Abstract Accurate delivery of low flow oxygen is essential for the management of infants with chronic respiratory failure. Twenty-seven examples of one low flow oxygen meter were tested. Measured flows were found to differ greatly from indicated settings, particularly at 250 mL/min settings. Knowledge of such errors as well as improved quality control are essential for appropriate management of oxygen therapy.
Keywords:low flow oxygen    oxygen flow meters