首页 | 本学科首页   官方微博 | 高级检索  
     


Cardiovascular magnetic resonance with an MR compatible pacemaker
Authors:Anita R Bhandiwad  Kristopher W Cummings  Michael Crowley  Pamela K Woodard
Affiliation:1.Division of Cardiology, Department of Medicine, Washington University School of Medicine, 1020 N. Mason Road, Suite 100, Saint Louis, MO, 63141, USA;2.Mallinckrodt Institute of Radiology, Washington University School of Medicine, Saint Louis, MO, USA
Abstract:Magnetic resonance imaging (MRI) within FDA guidelines for the MRI-conditional pacemaker precludes placing the heart at the center of the magnet’s bore. This in effect appears to preclude cardiovascular MR. In this manuscript, we describe a protocol for cardiovascular MR of patients with a Revo pacemaker system while operating within FDA guidelines, and the first US case of cardiovascular MR in a patient with a Revo MRI-conditional pacing system despite position constraints.
Keywords:MRI-conditional pacemaker   Cardiovascular magnetic resonance   Revo pacer