A New Calibration Method that Compensates for the Effects of O2 and N2 on Infrared CO2 Analysers |
| |
Authors: | Yongquan Tang M.B.B.S. Ph.D. Martin J. Turner Ph.D. MSc A. Barry Baker M.B.B.S. D.Phil. F.A.N.Z.C.A. F.R.C.A. F.J.F.I.C.M. |
| |
Affiliation: | (1) From the Department of Anaesthetics, University of Sydney, Royal Prince Alfred Hospital, Missenden Road, Camperdown, NSW 2050 Sydney, Australia |
| |
Abstract: | Objective. This study evaluates a method for calibrating mainstream CO2 analysers in which CO2 partial pressure (P CO2) is calculated as a function of the outputs of CO2 and O2 analysers. Methods. Three mass flow controllers were used to generate 25 different reference mixtures of O2, N2 and CO2. Reference gas mixtures were combinations of P CO2 = 2, 4, 6, 8, 10 kPa and O2 partial pressure (P O2) = 10, 20, 40, 60, 80 kPa (balance N2). CO2 and O2 analyser data were fitted by a calibration equation which took into account the effects of oxygen partial pressure and nonlinearity of the CO2 analyser. The calibration coefficients were tested in a separate validation data set with a variety of combinations of CO2 and O2. Results. Our new calibration method yields a standard deviation of CO2 measurement error that is significantly lower than a CO2-only calibration method in the validation data set (0.54% versus 2.72%, P < 0.05). P CO2 measurement errors produced by the single gas calibration equation are significantly correlated with P O2 in both the calibration (R = −0.9906, P < 0.05) and validation data sets (R = −0.9642, P < 0.05), but the errors given by our new calibration equation are independent of P O2 (R = −0.0364, NS, and R = −0.0305, NS, for calibration and validation data sets respectively). Calibration with only CO2 cannot eliminate the error related to the collision broadening effect of O2, which in our CO2 analyser is approximately a 1% underestimation of P CO2 for every 10 kPa (75 mmHg) increase in P O2. Conclusions. This study shows that non-dispersive infrared CO2 analyser readings can be substantially affected by background oxygen. This effect can be corrected for by calibrating the CO2 analyser with gases containing known proportions of both CO2 and O2. |
| |
Keywords: | infrared CO2 analyser calibration collision broadening |
本文献已被 PubMed SpringerLink 等数据库收录! |
|