首页 | 本学科首页   官方微博 | 高级检索  
     

基于无线能量传输技术的植入式动物生理参数遥测系统
引用本文:郭子平,刘华,颜国正,贾智伟,石煜,罗章源,朱柄全. 基于无线能量传输技术的植入式动物生理参数遥测系统[J]. 北京生物医学工程, 2012, 31(3): 288-292,301
作者姓名:郭子平  刘华  颜国正  贾智伟  石煜  罗章源  朱柄全
作者单位:上海交通大学电子信息与电气工程学院医学精密工程及智能系统研究所,上海,200240;杰升生物科技,上海,有限公司,上海,200235
摘    要:目的 为长期连续监测动物生理参数,设计并实现了一种基于无线能量传输技术的植入式遥测系统.方法 系统由无线能量传输设备、植入式微型电子胶囊、数据记录仪、数据处理软件组成.胶囊尺寸为13.06 mm×30.90 mm,重6.38 g,功耗约150 mW.结果 模拟实验验证了该植入式遥测系统能够监测心电、体温和血压.选用新西兰兔进行活体实验,测得体温范围为37.2~37.3℃,基线噪声为0.1℃,血压范围为74~83 mmHg,基线噪声为3 mmHg.结论 模拟实验和动物活体实验证明基于无线能量传输技术的植入式动物生理参数遥测系统能够比较准确地监测动物的心电、体温和血压等参数.

关 键 词:无线能量传输  植入式  遥测系统  生理参数

Implantable radio-telemetry system for monitoring animal physiological parameters based on wireless power transmission technology
GUO Ziping,LIU Hua,YAN Guozheng,JIA Zhiwei,SHI Yu,LUO Zhangyuan,ZHU Bingquan. Implantable radio-telemetry system for monitoring animal physiological parameters based on wireless power transmission technology[J]. Beijing Biomedical Engineering, 2012, 31(3): 288-292,301
Authors:GUO Ziping  LIU Hua  YAN Guozheng  JIA Zhiwei  SHI Yu  LUO Zhangyuan  ZHU Bingquan
Affiliation:1 Institute of Medical Precision Engineering and Intelligent System, School of Electronic Information and Electrical Engineering, Shanghai Jiao Tong University, Shanghai 200240 2 Genix Biotek Science Technology Company Limited, Shanghai 200235
Abstract:Objective An implantable radio-telemetry system based on wireless power transmission technology is designed for long-term monitoring animal physiological parameters continuously. Methods The system contains wireless power transmission device, implantable miniature electronic capsule, data logger and data processing software. The capsule is only ~b13.06 mm x 30.90 mm in size and weights 6.38 g with approximately 150 mW power consumption. Results The simulation experiment proves that the system is able to detect ECG,body temperature and blood pressure. One New Zealand rabbit is used in the vivo experiment,the detected body temperature is from 37.2 ~C to 37.3 ~C with 0. 1 ~C baseline noise and the detected blood pressure is from 74 mmHg to 83 mmHg with 3 mmHg baseline noise. Conclusions The proposed implantable radio-telemetry system, which is used for monitoring the animal physiological parameters based on wireless power transmission technology, is able to monitor the electrocardiograph (ECG), body temperature and blood pressure of animals.
Keywords:wireless power transmission  implantable  radio-telemetry system  physiologicalparameters
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号