首页 | 本学科首页   官方微博 | 高级检索  
     


Velocity fields and turbulent stresses downstream of biological and mechanical aortic valve prostheses implanted in pigs
Authors:J M Hasenkam  E M Pedersen  J H Ostergaard  H Nygaard  P K Paulsen  G Johannsen  B A Schurizek
Affiliation:Department of Thoracic and Cardiovascular Surgery, Aarhus Kommunehospital, Denmark.
Abstract:Since detailed knowledge about velocity fields downstream of heart valve prostheses obtained from in vitro studies has not been followed up by similar detailed studies in vivo a pig model for acute velocity field studies downstream of aortic valve prostheses was established. Two mechanical and two bioprosthetic valves were studied and a dynamic three dimensional visualisation of velocity fields one diameter downstream performed under different haemodynamic conditions in a total of 22 pigs. The Ionescu-Shiley pericardial valve had velocity fields very similar to the normal native porcine aortic valve. The Edwards-Carpentier porcine valve caused a jet type flow, and the valve design of the St Jude Medical and Bj?rk-Shiley Monostrut valves was reflected in the velocity profile. Normalised (mean(SEM] systolic Reynolds normal stresses in the total cross sectional area were: native porcine 15(1.5) Nm-2; St Jude Medical 24(3.4) Nm-2; Bj?rk-Shiley Monostrut 25(1.6) Nm-2; Edwards-Carpentier Supra-annular 51(6.6) Nm-2; Ionescu-Shiley Pericardial 19(2.0) Nm-2. Reynolds normal stresses were higher in areas of rapidly changing or constantly high velocity gradients.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏