Inappropriate Discharges of an Implantable Cardioverter Defibrillator Secondary to Automatic Adjustable Gain of Atrial Tachycardia
Authors:STUART O. SCHECTER  STEVEN M. GREENBERG  DAVID H. HOGH  MIGHAEL GESA  JOSEPH H. LEVINE
Affiliation:Cardiac Arrhythmia and Pacemaker Center, St. Francis Hospital., Roslyn, New York and North Suffolk Cardiology Associates, P.C., Stony Brook, New York
Abstract:Third generation implantable cardioverter defibrillators are capable of complex arrhythmia detection using sensing algorithms with automatic adjustable gain settings. We report a unique case where inappropriate sensing of atrial tachycardia in a patient with a His bundle ablation lead to satisfaction of ventricular fibrillation detection criteria.
Keywords:implantable cardioverter defibrillator    auto adjustable gain    oversensing    atrial tachycardia