首页 | 本学科首页   官方微博 | 高级检索  
检索        


Problems in Interpretation of Brainstem-Evoked Response Audiometry Results
Abstract:This report describes 6 cases - 3 adults and 3 children - in which no response could be obtained during the brainstem-evoked response audiometry (ABR) examination. This, in spite of the fact that maximal click intensity was used during the test, and despite the patients' nearly normal hearing sensitivity. All of the cases presented a slight degree of neurological deficiency, which could explain the faulty synchronization of the biological response to clicks. Attention is drawn to the fact that lack of response in ABR in cases with central nervous system involvement is not necessarily proof of severe hearing loss, and could supply important diagnostic cues.

L'article décrit 6 cas: ceux de 3 adultes et de 3 enfants chez qui le test ABR n'a donné aucun résultat, en dépit d'une intensité maximum du clic et malgré l'acuité auditive presque normale des malades. Certains des cas présentés ont montré de légères lésions neurologiques qui pourraient expliquer la synchronisation défectueuse de la réaction biologique au clic. La conclusion qui s'impose est que l'absence de réaction au test n'est pas nécessairement la preuve d'une perte auditive sévère ou d'une chute de l'audition dans les sons aigus, et que cela pourrait ětre important pour le diagnostic.
Keywords:Brainstem-evoked response audiometry examination  Interpretation of results  No response in brainstem-evoked response audiometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号