首页 | 本学科首页   官方微博 | 高级检索  
     


THE HALOSCALE "INFANTA" WRIGHT RESPIROMETER: An In Vitro and In Vivo Assessment
Authors:HATCH, D. J.   WILLIAMS, G. M. E.
Affiliation:The Hospital for Sick Children Great Ormond Street, London WCIN 3JH.
Ferraris Development & Engineering Ltd 26 Lea Valley Trading Estate, Angel Road, Edmonton, London N18 3JD
Abstract:The performance of the Haloscale "Infanta" respirometer hasbeen assessed in vitro using ISO test compliances and resistances,and in vivo by comparison with pneumotachograph volumes in 13spontaneously breathing children and 13 children during intermittentpositive pressure ventilation. The Infanta was shown to be capableof registering volumes between 15 and 200 ml with an accuracyof ±5%. The registered volume decreased rapidly below15 ml, whilst above 200 ml over-registration developed.
Keywords:
本文献已被 Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏