Transapical Transcatheter Aortic Valve for Severe Aortic Regurgitation: Expanding the Limits |
| |
Affiliation: | 1. Department of Thoracic and Cardiovascular Surgery, West-German Heart Center Essen, University Hospital Essen, Germany;2. Department of Cardiology, West-German Heart Center Essen, University Hospital Essen, Essen, Germany |
| |
Abstract: | ObjectivesThis study sought to evaluate the self-expandable ACURATE TA device (Symetis SA, Ecublens, Switzerland) in a cohort of patients with pure aortic regurgitation (AR).BackgroundTranscatheter aortic valve replacement (TAVR) has been initially considered as an alternative for high-risk patients with aortic stenosis. Although the current experience is limited, TAVR might be also an alternative to treat patients with pure, severe AR.MethodsBetween April 2012 and December 2013, a total of 8 high-risk patients with pure, severe AR were enrolled (grade III+). Clinical and hemodynamic data as well as data on device and procedure parameters and outcomes were collected.ResultsPatient mean was 72.5 ± 8.4 years, and 37.5% of patients were female. Logistic EuroSCORE was 34.0 ± 7.9% and the Society of Thoracic Surgeons score was 7.3 ± 3.3% on average. Two patients had undergone emergency aortic operation before due to acute type A aortic dissection, and both were treated by replacement of the ascending aorta (including root reconstruction) and the aortic arch combined with or without E-vita Open stent graft (Jotec GmbH, Hechingen, Germany) (January 2011 and March 2012), whereas the other patients experienced primary AR. All patients underwent successful transapical TAVR with the transapical ACURATE TA device (size small, n = 1, size medium, n = 3, size large, n = 4) without any intraprocedural complications according to the Valve Academic Research Consortium 2 criteria. Post-procedure AR grade I+ or lower, as revealed by transoesophageal echocardiography and angiography, was present in all 8 patients. At 30 days, the stroke incidence and all-cause mortality rate were 0%.ConclusionsThis small single-center series demonstrates the feasibility of transapical TAVR with the self-expandable ACURATE TA device in high-risk patients with severe AR. |
| |
Keywords: | aortic regurgitation aortic valve high-risk transcatheter aortic valve replacement AAD" },{" #name" :" keyword" ," $" :{" id" :" kwrd0035" }," $$" :[{" #name" :" text" ," _" :" acute aortic dissection AR" },{" #name" :" keyword" ," $" :{" id" :" kwrd0045" }," $$" :[{" #name" :" text" ," _" :" aortic regurgitation NYHA" },{" #name" :" keyword" ," $" :{" id" :" kwrd0055" }," $$" :[{" #name" :" text" ," _" :" New York Heart Association TAVR" },{" #name" :" keyword" ," $" :{" id" :" kwrd0065" }," $$" :[{" #name" :" text" ," _" :" transcatheter aortic valve replacement THV" },{" #name" :" keyword" ," $" :{" id" :" kwrd0075" }," $$" :[{" #name" :" text" ," _" :" transcatheter heart valve TEE" },{" #name" :" keyword" ," $" :{" id" :" kwrd0085" }," $$" :[{" #name" :" text" ," _" :" transesophageal echocardiography TTE" },{" #name" :" keyword" ," $" :{" id" :" kwrd0095" }," $$" :[{" #name" :" text" ," _" :" transthoracic echocardiography VARC" },{" #name" :" keyword" ," $" :{" id" :" kwrd0105" }," $$" :[{" #name" :" text" ," _" :" Valve Academic Research Consortium |
本文献已被 ScienceDirect 等数据库收录! |
|