Comparisons of 17 lots of 2-oxoglutarate, and specifications for use of this substrate in reference methods
Authors:P N Bowers  G N Bowers  R B McComb
Abstract:We examined 17 lots of 2-oxoglutarate (seven acid forms, three K salt forms, and seven Na salt forms), obtained from eight commercial suppliers, for suitability for measuring aspartate aminotransferase (EC 2.6.1.1) and alanine aminotransferase (EC 2.6.1.2) in human serum. Measurements of the catalytic activity concentrations of these two aminotransferases with each of these 17 preparations were not sufficiently sensitive to distinguish good from poor-quality material. Thus, we ranked these lots for purity, by specific analysis with glutamate dehydrogenase and by liquid chromatography, and determined the water content, acid content, and spectral characteristics of each. On the basis of a 2-oxoglutarate assay value by glutamate dehydrogenase of 98% or greater, we considered seven of the preparations acceptable and 10 unacceptable. The molar absorptivities (L X mol-1 X cm-1, mean +/- SD) of the seven acceptable lots in 1 mol/L HCl were: epsilon 325 nm = 9.12 +/- 0.02 (CV = 0.2%), epsilon 279 nm = 2.63 +/- 0.23 (CV = 9.9%), and epsilon 245 nm = 37.9 +/- 4.1 (CV = 10.9%). Use of these spectrophotometric limits alone unambiguously distinguished the inferior lots of 2-oxoglutarate. We urge the inclusion of detailed spectrophotometric specifications for 2-oxoglutarate in Reference Methods for aminotransferase measurements.
Keywords: