首页 | 本学科首页   官方微博 | 高级检索  
检索        

全膝关节置换后关节线变化对髌股关节生物力学的影响
作者姓名:  
作者单位:河南中医学院第一附属医院骨伤科,河南省郑州市 450000
摘    要:

关 键 词:骨科植入物  人工假体  全膝关节置换  关节线  假体  聚乙烯衬垫  三维有限元  髌股关节  生物力学  
收稿时间:2016-02-03

Effects of joint line changes on biomechanics of patellofemoral joint after total knee arthroplasty
Authors:Wang Tao
Institution:Department of Orthopedics and Traumatology, First Affiliated Hospital, Henan University of Chinese Medicine, Zhengzhou 450000, Henan Province, China
Abstract:BACKGROUND: In addition to infection, the reasons for total knee arthroplasty revision are polyethylene liner wear and prosthesis loosening. The impact of joint line height on patellofemoral biomechanics was inconclusive. OBJECTIVE: To study the impact of joint line changes after total knee arthroplasty on patellofemoral joint biomechanics. METHODS: Three-dimensional finite element models were established after total knee arthroplasty. The impact of different joint line height on quadriceps tensile force, patella tendon tension, and patellofemoral joint forces was calculated at range of flexion of 0°, 30°, 60° and 90°.  RESULTS AND CONCLUSION: (1) At knee flexion of 0°, the height of the joint line in -3 mm-4.5 mm did not have great impacts on quadriceps tensile force, patella tendon tension and patellofemoral joint forces. (2) At knee flexion of 30°-90°, the height of the joint line below 3 mm did not have great impacts on quadriceps tensile force, patella tendon tension and patellofemoral joint forces. The height of the joint line more than 3 mm had obvious impacts on quadriceps tensile force, patella tendon tension and patellofemoral joint forces. (3) It was recommended that the height of joint line in total knee arthroplasty was preferably controlled within 3 mm. 
Keywords:
点击此处可从《》浏览原始摘要信息
点击此处可从《》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号