首页 | 本学科首页   官方微博 | 高级检索  
     

金宝AK 200 S血透机气密测试故障维修分析
引用本文:牵金宝. 金宝AK 200 S血透机气密测试故障维修分析[J]. 中国卫生产业, 2013, 0(23): 180-181
作者姓名:牵金宝
作者单位:大庆油田总医院,黑龙江大庆163411
摘    要:本文简单介绍了AK200S血液透析机气密测试自检的分析,通过对故障案例的维修分析与总结,旨在针对不同气密故障类型分别判断出具体故障部位。

关 键 词:血液透析机维修  气密测试  FRN技术  分段打压法

Troubleshooting of the leak test for the Gambro AK 200ULTRA S dialysis machine
LI Jinbao. Troubleshooting of the leak test for the Gambro AK 200ULTRA S dialysis machine[J]. China Health Industry, 2013, 0(23): 180-181
Authors:LI Jinbao
Affiliation:LI Jinbao( General Hospital of Daqing Oil Field, Heilongjiang 163411, China)
Abstract:This article mainly introduces the Gambro AK 200ULTRA S dialysis machine during the self check on leak test, analyzes the causes of the malfunction and puts forward the measures in order to determine the specific fault location for different fault types of airtight.
Keywords:Dialysis machine maintenance  Leak test  FRN technology  Sectional pressure method
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号