首页 | 本学科首页   官方微博 | 高级检索  
     


Estimating T1 from multichannel variable flip angle SPGR sequences
Authors:Joshua D. Trzasko  Petrice M. Mostardi  Stephen J. Riederer  Armando Manduca
Affiliation:1. Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, Minnesota, USA;2. Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA
Abstract:Quantitative estimation of T1 is a challenging but important task inherent to many clinical applications. The most commonly used paradigm for estimating T1 in vivo involves performing a sequence of spoiled gradient‐recalled echo acquisitions at different flip angles, followed by fitting of an exponential model to the data. Although there has been substantial work comparing different fitting methods, there has been little discussion on how these methods should be applied for data acquired using multichannel receivers. In this note, we demonstrate that the manner in which multichannel data is handled can have a substantial impact on T1 estimation performance and should be considered equally as important as choice of flip angles or fitting strategy. Magn Reson Med, 2013. © 2012 Wiley Periodicals, Inc.
Keywords:T1 mapping  relaxometry  variable flip angle  non‐linear least squares
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号