Abdomen: diffusion-weighted MR imaging with pulse-triggered single-shot sequences |
| |
Authors: | Mürtz Petra Flacke Sebastian Träber Frank van den Brink Johan S Gieseke Jürgen Schild Hans H |
| |
Institution: | Department of Radiology, University of Bonn, Sigmund-Freud-Strasse 25, D-53105 Bonn, Germany. muertz@uni-bonn.de |
| |
Abstract: | Magnetic resonance (MR) diffusion measurements of the abdomen were performed in 12 healthy volunteers by using a diffusion-weighted single-shot sequence both without and with pulse triggering for different trigger delays. Pulse triggering to the diastolic heart phase led to reduced motion artifacts on the diffusion-weighted MR images and to significantly improved accuracy and reproducibility of measurements of the apparent diffusion coefficients, or ADCs, of abdominal organs. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
| 点击此处可从《Radiology》浏览原始摘要信息 |