An extension set for 81Rb-81mKr solution generators for lung ventilation studies
Affiliation:1. Chemistry Division, Los Alamos National Laboratory, P.O. Box 1663, Los Alamos, NM 87545, USA;2. Collider-Accelerator Department, Brookhaven National Laboratory, Bldg 801, Upton, NY 11973, USA
Abstract:For 81Rb-81mKr solution generators, an extension set has been developed, which strips the 81mKr from the liquid eluate with a stream of air for use in lung ventilation studies. Via a three-way valve the 81Rb-81mKr solution generator can be operated alternately in the perfusion mode or in the ventilation mode. Measurements and calculations have been performed to get more insight into the parameters of interest for an optimum design of the extension set and its operation conditions.
Keywords:
本文献已被 ScienceDirect 等数据库收录!