Detectability of simulated pulmonary nodules on chest radiographs: Comparison between irradiation side sampling indirect flat-panel detector and computed radiography
Authors:Yuki Yano  Hidetake Yabuuchi  Nobukazu Tanaka  Junji Morishita  Tsutomu Akasaka  Yoshio Matsuo  Shunya Sunami  Takeshi Kamitani  Mikako Jinnouchi  Yuzo Yamasaki  Michinobu Nagao  Masayuki Sasaki
Institution:1. Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka 812-8582, Japan;2. Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka 812-8582, Japan;3. Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka 812-8582, Japan
Abstract:

Objective

To compare the detectability of simulated pulmonary nodules on chest radiographs between an irradiation side sampling indirect flat-panel detector (ISS-FPD) and computed radiography (CR).

Materials and methods

This study was an observer performance study. Simulated pulmonary nodules of 8 mm in diameter were superimposed on an anthropomorphic chest phantom. Chest radiographs were acquired under 2 exposure levels (4 and 3.2 mAs) with the ISS-FPD and the CR. Six thoracic radiologists evaluated all 40 images (10 patterns × 2 different exposure doses × 2 different systems) for the presence or absence of a lesion over each of 12 defined areas on a 3-megapixel monochrome liquid-crystal display. Receiver operating characteristic (ROC) curves were obtained for observation in predefined 480 areas. A jackknife method was used for statistical analysis. Differences with a P value of <0.05 were considered significant.

Results

The analysis of the observer detection of simulated pulmonary nodules showed larger areas under the ROC curve (AUC) by the ISS-FPD than by the CR. There was a statistically significant difference between the two systems at 3.2 mAs (P = 0.0330).

Conclusion

The ISS-FPD was superior to the CR for the detection of simulated pulmonary nodules at 3.2 mAs.
Keywords:Radiographic image enhancement  Solitary pulmonary nodule  Phantoms  Imaging  Diagnostic performance  ROC curve
本文献已被 ScienceDirect 等数据库收录!