kavo牙科综合治疗机的技术改造
引用本文:胡轶琳,陈增喜. kavo牙科综合治疗机的技术改造[J]. 医疗卫生装备, 2010, 31(7): 66-67
作者姓名:胡轶琳  陈增喜
作者单位:大庆市第五医院设备科,黑龙江大庆,163711
摘    要:目的:对kavo牙科综合治疗机进行技术改造,以解决气动组合进水阀的漏水问题。方法:经过大量的比对,采用经过特殊处理的螺栓代替气动进水阀的阀芯。结果:器件替换后解决了牙科治疗椅的漏水问题。结论:该方法不仅节省了大量资金,而且缩短了等待配件的时间,变相地节约了成本。

关 键 词:牙科综合治疗机  气动进水阀  阀芯  螺栓

Technical Improvement for kavo Dental Multipurpose Therapy Unit
HU Yi-lin,CHEN Zeng-xi. Technical Improvement for kavo Dental Multipurpose Therapy Unit[J]. Chinese Medical Equipment Journal, 2010, 31(7): 66-67
Authors:HU Yi-lin  CHEN Zeng-xi
Affiliation:(Equipment Department,the Fifth Hospital of Daqing,Daqing,Heilongjiang Province 163711,China)
Abstract:Objective To solve the leakage problem of air inlet valve of kavo dental multipurpose therapy unit.Methods The bad air inlet valve was replaced by the special bolt.Results The leakage problem of air inlet valve was solved.Conclusion The technical improvement of kavo dental multipurpose therapy unit results in decreased cost and waiting time for the spare parts.
Keywords:dental multipurpose therapy unit  air inlet valve  valve disc  bolt