医用观片灯的现状调查及分析
引用本文:王先运 杜冰. 医用观片灯的现状调查及分析[J]. 中国医疗器械杂志, 1995, 19(5): 282-283,297
作者姓名:王先运 杜冰
作者单位:泰山医学院放射系,山东省医学影象研究所
摘    要:笔者对医用观片灯的现状进行了调查,结果发现被检的66台灯中,最高亮度1305Lx,最低160Lx,平均760Lx,达不到国家卫生部的标准。亮度严重不足,直接影响了诊断质量,亟待解决。

关 键 词:观片灯 亮度 光度测定法 X线照片 X线诊断

Investigation and analysis of medical viewboxs
Wang Xianyun,Li Yueqing,Zhang ZhaoChen,Ding Congfu. Investigation and analysis of medical viewboxs[J]. Chinese journal of medical instrumentation, 1995, 19(5): 282-283,297
Authors:Wang Xianyun  Li Yueqing  Zhang ZhaoChen  Ding Congfu
Affiliation:Wang Xianyun;Li Yueqing;Zhang ZhaoChen;Ding Congfu (Department of Radiology Taishan Medical College)Du bing;Ding Shaojuan (Shandong Medical Imaging Research Institute)
Abstract:After we investigated the condition of 66 viewboxs used in hospitals, we discovered that none of them are up to the standard. The maximum brightness of them is 1305 Lx, the minimum is 160LX, and the average is 760 Lx. It is far lower then the standard of Ministry of Health (the brightness of viewbox must not be lower than 1350 cd/m2, i.e. 1350 Lx). The brightness of the viewboxes is so insuff icient that it has directly had influence on the diagnosis quality.For this reason. it must be corrected imme diately.
Keywords:viewbox  brightness  photometry  radiogram  