New devices for imaging in nuclear medicine
Authors:Pani Roberto  Pellegrini Rosanna  Cinti Maria Nerina  Trotta Carlo  Bennati Paolo  Betti Margherita  De Vincentis Giuseppe  Cusanno Francesco  Garibaldi Franco  Ridolfi Stefano  Majewsky Stan  Tsui Benjamin M W
Affiliation:Department of Experimental Medicine and Pathology, University "La Sapienza," Rome, Italy. roberto.pani@uniroma1.it
Abstract:Pinhole gamma camera imaging offers the ability to obtain high resolution images from single gamma ray emitting radiotracers playing a reasonable tradeoff between very small field of view (FoV) and sensitivity. On the other hand the total spatial resolution of a pinhole imaging device is predominantly affected by the detector intrinsic spatial resolution for reduced magnification factors. To design very compact pinhole SPET scanners with very high intrinsic spatial resolution, authors investigated a miniature gamma camera based on the newly developed Hamamatsu H8500 flat panel photomultiplier. The PSPMT was coupled to the following scintillation arrays: CsI(Tl) array with 0.2-mm, 1-mm, 1.4-mm pixel size and NaI (Tl) with 1-mm pixel size. The imaging performances were evaluated by 57Co spot and flood irradiations. NaI(Tl) array shows a better pixel identification for 1 mm pixel size, proving to be a good candidate to make a large area photodetector based on multi PSPMTs closely packed. Although CsI(Tl) array had the smallest pixel size, the low light output limited the best intrinsic spatial resolution to about 0.5 mm.
Keywords:
本文献已被 PubMed 等数据库收录!