多层面螺旋CT肝脏双期增强技术的研究—造影剂量及团注造影剂跟踪技术的影响
引用本文:周良平,郭青,徐香玖,沈国强. 多层面螺旋CT肝脏双期增强技术的研究—造影剂量及团注造影剂跟踪技术的影响[J]. 中国临床医学影像杂志, 2001, 12(6): 416-418
作者姓名:周良平  郭青  徐香玖  沈国强
作者单位:甘肃省人民医院CT室,甘肃,兰州,730000
摘    要:目的:评价造影剂量及团注造影剂跟踪技术对多层面螺旋CT肝脏双期增强效果的影响。材料与方法:40名观察对象随机分为三组,分别从周围静脉注入Omnipaque 300 100ml、100ml和75ml。第一组以固定时间间隔扫描,第二组及第三组以团注造影剂跟踪技术启动扫描。以腹主动脉增强值超过150Hu为动脉期开始,肝实质增强值大于20Hu为动脉期结束。结果:①第一组中5例(45%)动脉期扫描序列未处于其最佳扫描间期,第二组为2例(13%),第三组为1例(8%);②第一组中2例(18%)肝实质强化峰值未包括在其门静脉期扫描序列中,2例(18%)强化峰值位于扫描序列的末期,而第二组和第三组的强化峰值均位于其扫描序列的中心层面;③各组间主动脉及肝实质增强值的差异无统计学意义。结论:①团注造影剂跟踪技术能更准确确定动脉期的起点;②75ml 的造影剂量能够满足多层面螺旋CT肝脏双期增强的要求。

关 键 词:肝疾病 造影剂 螺旋CT
文章编号:1008-1062(2001)06-0416-04
修稿时间:2001-05-17

Dual phase multi slice helical CT of the liver:effect of contrast material volume and bolus tracking
ZHOU Liang ping,GUO Qing,XU Xiang jiu,SHEN Guo qiangCT Division,People's Hospital of Gansu Province,Lanzhou ,China. Dual phase multi slice helical CT of the liver:effect of contrast material volume and bolus tracking[J]. Journal of China Clinic Medical Imaging, 2001, 12(6): 416-418
Authors:ZHOU Liang ping  GUO Qing  XU Xiang jiu  SHEN Guo qiangCT Division  People's Hospital of Gansu Province  Lanzhou   China
Affiliation:ZHOU Liang ping,GUO Qing,XU Xiang jiu,SHEN Guo qiangCT Division,People's Hospital of Gansu Province,Lanzhou 730000,China
Abstract:Objective:To assess the effects of contrast material volume and bolus tracking in dual phase multi slice helical CT of the liver.Materials and Methods:Fourty patients were randomly divided into three groups with different volumes of Omnipaque300:100ml,100ml and 75ml.In group one,the scanning delays were fixed,whereas in group two and three,scanning delays were determined individually by means of a semiautomatic bolus tracking device.The optimal scanning intervals(OSI) of arterial phase for MSCT were defined as the interval when the enhancement value for the aorta reached 150Hu and ending when the value for the liver parenchyma reached 20Hu.Results:1In group one,the arterial scanning sequences of five cases were not within their OSI,but in group two and three there were only two cases and one case;2During the portal venous phase,the liver parenchyma peak enhancement of two cases was not within the portal venous scanning sequences(PVSS) and that of two cases were near the ending of their PVSS in group one,whereas in group two and three,the peak enhancement of liver parenchyma was all near the center of their PVSS.3No significant difference of the peak enhancement of aorta and liver parenchyma were observed among these three groups.Conclusion:1A semiautomatic bolus tracking device can determine the OSI more precisely in dual phase MSCT of the liver;2For MSCT contrast medium volume can be decreased to 75ml effectively.
Keywords:liver diseases  tomography  X-ray computed  contrast media  