首页 | 本学科首页   官方微博 | 高级检索  
     


Bronchoscopy via a Redesigned Combitube(TM) in the Esophageal Position: A Clinical Evaluation
Authors:Krafft, Peter MD   Roggla, Martin MD   Fridrich, Peter MD   Locker, Gottfried J. MD   Frass, Michael MD   Benumof, Jonathan L. MD
Abstract:Background: The esophageal-tracheal Combitube (Kendall-Sheridan Catheter Corp., Argyle, NY) is an effective device for providing adequate gas exchange. However, tracheal suctioning is impossible with the Combitube placed in the esophageal position. To eliminate this disadvantage, the Combitube was redesigned by creating an enlarged hole in the pharyngeal lumen that allows fiberoptic access, tracheal suctioning, and tube exchange over a guide wire.

Methods: The two anterior, proximal perforations of regular Combitubes were replaced by a larger, ellipsoid-shaped hole. After the study was approved by the institutional review board, 20 patients with normal airways (Mallampati I or II) were studied. During general anesthesia, patients were esophageally intubated with the Combitube. A flexible bronchoscope was inserted and guided via the modified hole and glottic opening down the trachea. For the replacement procedure, a J tip guide wire was introduced through the bronchoscope. The bronchoscope and the Combitube were removed and a standard endotracheal tube was advanced over a guide catheter.

Results: Bronchoscopic evaluation of the trachea and guided replacement of the Combitube by an endotracheal tube was successful in all 20 study patients. The average time needed to perform airway exchange was 90 +/- 20 s (mean +/- SD). Arterial oxygen saturation and end-tidal carbon dioxide levels remained normal in all patients. No case of laryngeal trauma was observed during intubation or the airway exchange procedure.

Keywords:
点击此处可从《Anesthesiology》浏览原始摘要信息
设为首页 | 免责声明 | 关于勤云 | 加入收藏