首页 | 本学科首页   官方微博 | 高级检索  
     

多模型迭代重建在CT门静脉成像中改善图像质量的应用
引用本文:任占丽1,2,贺太平2,于勇2,于楠2,3,张喜荣2,3,贾永军2,段海峰2,李豆4,贺朝1,彭雪艳5. 多模型迭代重建在CT门静脉成像中改善图像质量的应用[J]. 中国医学物理学杂志, 2020, 37(8): 1017-1021. DOI: DOI:10.3969/j.issn.1005-202X.2020.08.015
作者姓名:任占丽1  2  贺太平2  于勇2  于楠2  3  张喜荣2  3  贾永军2  段海峰2  李豆4  贺朝1  彭雪艳5
作者单位:1.陕西中医药大学第二附属医院医学影像中心, 陕西 咸阳 712000; 2.陕西中医药大学附属医院医学影像科, 陕西 咸阳 712000; 3.陕西中医药大学医学技术学院, 陕西 咸阳 712000; 4.长安医院医学影像科, 陕西 西安 710016; 5.咸阳市中心医院, 陕西 咸阳 712000
摘    要:目的:探讨多模型迭代重建(ASIR-V)在CT门静脉成像(CTPV)中改善图像质量的应用价值。方法:收集Revolution CT检查CTPV患者60例并随机分为A、B组。两组均采用管电压120 kV、噪声指数10、对比剂用量450 mgI/kg。扫描结束后,A组采用40%ASIR,B组采用40%ASIR-V。分别测量门静脉主干、门静脉左支、门静脉右支的CT值和CT值标准差,并计算信噪比和对比噪声比。记录两组患者CT容积剂量指数(CTDIVOL)、剂量长度乘积(DLP),并计算有效辐射剂量(ED)。由两名有经验放射科医师采用5分法标准对图像质量进行双盲法主观评分。结果:两组患者一般资料、辐射剂量(CTDIVOL、DLP、ED)和对比剂总量间差异无统计学意义(P>0.05)。两组门静脉主干、门静脉左支及右支CT值均无统计学差异(P>0.05)。A组门静脉主干、门静脉左支及右支CT值的标准差高于B组,A组信噪比和对比噪声比低于B组,且差异均具有统计学意义(P<0.05)。两名医师对两组图像主观评分具有很好的一致性(Kappa值>0.75, P<0.05),B组图像主观评分显著高于A组(P<0.05)。结论:ASIR-V可以显著提高门静脉图像质量,为临床诊断提供更好的CTPV图像。

关 键 词:CT门静脉成像  多模型迭代重建  图像质量

Application of adaptive statistical iterative reconstruction-V technique in CT portal venography for improving image quality
REN Zhanli1,2,HE Taiping2,YU Yong2,YU Nan2,3,ZHANG Xirong2,3,JIA Yongjun2,DUAN Haifeng2,LI Dou4,HE Chao1,PENG Xueyan5. Application of adaptive statistical iterative reconstruction-V technique in CT portal venography for improving image quality[J]. Chinese Journal of Medical Physics, 2020, 37(8): 1017-1021. DOI: DOI:10.3969/j.issn.1005-202X.2020.08.015
Authors:REN Zhanli1  2  HE Taiping2  YU Yong2  YU Nan2  3  ZHANG Xirong2  3  JIA Yongjun2  DUAN Haifeng2  LI Dou4  HE Chao1  PENG Xueyan5
Affiliation:1. Department of Radiology, the Second Affiliated Hospital of Shaanxi University of Chinese Medicine, Xianyang 712000, China 2. Department of Radiology, Affiliated Hospital of Shaanxi University of Chinese Medicine, Xianyang 712000, China 3. College of Medical Technology, Shaanxi University of Chinese Medicine, Xianyang 712000, China 4. Department of Radiology, Changan Hospital, Xian 710016, China 5. Xianyang Central Hospital, Xianyang 712000, China
Abstract:Abstract: Objective To explore the application value of adaptive statistical iterative reconstruction-V (ASIR-V) technique in improving the quality of computed tomography portal venography (CTPV) image. Methods Sixty patients who underwent CTPV using Revolution CT were enrolled and randomly divided into group A and group B. Tube voltage of 120 kV, noise index of 10 and contrast agent dose of 450 mgI/kg were adopted in both two groups. After scanning, group A was reconstructed with 40%ASIR, while group B was reconstructed with 40%ASIR-V. The CT values and its standard deviations of the main portal vein (MPV), the left branch of portal vein (LPV), and the right branch of portal vein (RPV) were measured, and their signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated. The volume CT dose index (CTDIVOL) and dose length product (DLP) in two groups were recorded, and the effective dose (ED) was calculated. The subjective scores of image quality were assessed blindly by two experienced radiologists using a 5-point system. Results There was no significant difference between two groups in general information, irradiation dose (CTDIVOL, DLP, ED) and the total contrast agent dose (P>0.05). The CT values of MPV, LPV and RPV showed no statistical difference between two groups (P>0.05). Moreover, compared with group B, group A had higher standard deviations of CT values and lower SNR and CNR for MPV, LPV and RPV, with statistical significance (P<0.05). There was a good agreement between two radiologists (Kappa>0.75, P<0.05). The subjective score of the images in group B was significantly higher than that in group A (P>0.05). Conclusion ASIR-V technique can significantly improve the quality of CTPV image, thereby providing better CTPV image for clinical diagnosis.
Keywords:Keywords: computed tomography portal venography adaptive statistical iterative reconstruction-V image quality
本文献已被 CNKI 等数据库收录!
点击此处可从《中国医学物理学杂志》浏览原始摘要信息
点击此处可从《中国医学物理学杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号