首页 | 本学科首页   官方微博 | 高级检索  
     


Low dose rate teletherapy using a telecaesium 137 unit radiobiological, physical and clinical considerations
Authors:S G Cooper  A P Cardew  S Ferguson  D J Joseph  C S Hamilton  J W Denham  A R Williams
Affiliation:Department of Radiation Oncology, Newcastle Mater Misericordiae Hospital.
Abstract:Low dose rate teletherapy aims to combine the supposedly superior results obtained with low dose rate implants with the convenience and staff protection characteristics of external beam therapy. Previous investigators have used telecobalt units to produce dose rates of 1.1 to 1.8 Gy/hr to treat in daily sessions lasting 6-10 hours to total doses of 60-70 Gy. These studies have not discounted the possibility that much of the advantage of interstitial implants results from the low dose rates used per se, and from the fact that the total dose is delivered in a short overall time. The relationship between total dose, dose rate and volume giving normal tissue and anit-tumour effects, however, remains ill-defined. At the Newcastle Mater Misericordiae Hospital a Caesium teletherapy unit has been modified to treat at low dose rates and a study has been designed with a view to establish which permutations of total dose and dose rate are isoeffective for acute mucosal and acute skin reactions in the dose rate range between 0.8 and 9.6 Gy/hr (1.3 and 16 cGy/min).
Keywords: