首页 | 本学科首页   官方微博 | 高级检索  
检索        


Evaluation of a hand-held spirometer, the Respiradyne, for the measurement of forced expiratory volume in the first second (FEV1), forced vital capacity (FVC) and peak expiratory flow rate (PEFR)
Authors:S C Jenkins  N C Barnes  J Moxham
Institution:Department of Thoracic Medicine, King's College School of Medicine and Dentistry, London.
Abstract:A hand-held spirometer, the Respiradyne (R), was evaluated for the measurement of forced expiratory volume in the first second (FEV1), forced vital capacity (FVC) and peak expiratory flow rate (PEFR) by comparing it with a Vitalograph spirometer (V) and a Wright's peak flow meter (W) in 70 subjects (normals and patients). The results showed close agreement for FEV1; r = 0.99, R = 0.961V + 0.03 X 10(-5) and FVC; r = 0.99, R = 1.003V-0.044. Results for PEFR using the Respiradyne were generally higher than with the peak flow meter; r = 0.98, R = 1.151W-17.576. The Respiradyne is portable and simple to operate and may be suited to use in a variety of non-laboratory situations.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号