The Activitrax rate responsive pacemaker system
Authors:K den Dulk  L Bouwels  F Lindemans  I Rankin  P Brugada  H J Wellens
Affiliation:Department of Cardiology, Academic Hospital Meastricht, University of Limburg, The Netherlands.
Abstract:Bipolar Medtronic Activitrax rate responsive pacemakers were implanted in 31 patients for ventricular (28) or atrial (3) pacing. Mean follow-up was 16 months (range 10 to 26). Twenty pacemakers were implanted after catheter ablation of the His bundle, 7 for sick sinus syndrome. 1 for atrioventricular block and 3 for sick sinus syndrome with atrioventricular block. A rate response value was selected that gave a pacing rate of about 100 pulses/min during walking. Of the 31 patients, all had 24-hour ambulatory electrocardiographic monitoring with diary, 11 walked a 20-minute circuit, including a flight of stairs, and 20 had a treadmill exercise test. In 9 patients the pacing rate could be compared with the underlying sinus rate during exercise and was seen to match it very closely. In 12 patients the pacing rate during car driving was found to be similar to the sinus rate of 5 volunteers under similar conditions (mean minimum and maximum rate was 80 and 99 pulses/min, respectively). No pacing-induced arrhythmias were seen during ambulatory electrocardiographic monitoring. At high pacing rates slightly irregular pacing intervals were sometimes observed, which was due to polarization sensing. Sporadically, 1 pacing interval shortened to the upper rate value, because of a known and now resolved timing anomaly. Neither anomaly was of clinical consequence and the first could be resolved by reprogramming.(ABSTRACT TRUNCATED AT 250 WORDS)
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号