Development of a small wireless device for perspiration monitoring
Affiliation:1. Wellness Promotion Science Center, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942, Japan;2. Advanced Research Center for Human Sciences, Waseda University, 2-579-15 Mikajima, Tokorozawa, Saitama 359-1192, Japan;3. Department of Clinical Laboratory Science, Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942, Japan;4. Asanogawa General Hospital, Naka 83, Kosaka-machi, Kanazawa, Ishikawa 920-8621, Japan;1. Institute of Environmental Radioactivity, Fukushima University, 1 Kanayagawa, Fukushima City, Fukushima 960-1296, Japan;2. Department of Civil Engineering, Southern University, 739/A Mehedibag Road, Chittagong 4000, Bangladesh;3. Institute of Science and Engineering, Kanazawa University, Kakuma, Kanazawa 920-1192, Japan;1. Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 9200942, Japan;2. Department of Central Radiology, Nagoya City University Hospital, 1-Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 4678602, Japan;3. Department of Radiology, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa 9208641, Japan
Abstract:A small and wireless device that can capture the temporal pattern of perspiration by a novel structure of water vapor collection combined with reusable desiccant has been developed. The novel device consists of a small cylindrical case with a temperature/relative humidity sensor, battery-driven data logger, and silica gel (desiccant). Water vapor of perspiration was detected by the change in relative humidity and then adsorbed by silica gel, allowing continuous recording of perspiration within a closed and wireless chamber, which has not been previously achieved. By comparative experiments using the commercially-available perspiration monitoring device, the developed device could measure perspiration as efficiently as the conventional one, with a normalized cross coefficient of 0.738 with 6 s delay and the interclass correlation coefficient [ICC(2, 1)] of 0.84. These results imply a good agreement between the conventional and developed devices, and thus suggest the applicability of the developed device for perspiration monitoring.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号