首页 | 本学科首页   官方微博 | 高级检索  
     


An isocentrically mounted stand for total body irradiation
Authors:Jones D  Rieke J W  Madsen B L  Hafermann M D
Affiliation:Northwest Medical Physics Center, Lynnwood, WA, USA.
Abstract:The purpose of this study was to construct a stand to support a patient for total body photon irradiation and to expedite the set-up and treatment by rotating the stand. As in other isocentric treatments, the midline dose is impacted less by source-to-skin distance variations. The method of immobilizing the patient is described. A 10 mm lucite plate is supported in front of the patient to increase skin dose. A matrix of holes in this plate serves to index the location of blocks used to shield the lungs. The dosimetry of the set-up is described, as is the production of tissue deficit compensators. The results of phantom studies and in vivo thermoluminescent dosimetry measurements are presented.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏