64排螺旋CT颈部血管成像低电压扫描初步研究
引用本文:郑少燕,初建平,杨智云,严超贵,范淼,周旭辉,李树荣.64排螺旋CT颈部血管成像低电压扫描初步研究[J].影像诊断与介入放射学,2011,20(5):323-326.
作者姓名:郑少燕  初建平  杨智云  严超贵  范淼  周旭辉  李树荣
作者单位:1. 汕头市中心医院放射科,广东,515031
2. 中山大学附属第一医院放射科,广州,510080
基金项目:广东省自然科学基金项目,广东省科技计划项目,教育部博士点基金新教师项目
摘    要:目的 探讨低管电压(100kV)扫描在颈部CT血管成像(CTA)的临床应用可行性.方法 将55名临床需要做颈部CTA检查的患者(其中男37例,女18例,平均54岁)按随机分配原则编入A、B两组(A组27例,B组28例),分别采用管电压100及120kV,管电流均采用150mAs.比较两组的CT图像质量(包括动脉平均强化...

关 键 词:体层摄影术  X线计算机  低剂量  CT血管成像  辐射剂量

Study of 64 slice detector carotid CT angiography with low tube voltage
ZHENG Shao-yan,CHU Jian-ping,YANG Zhi-yun,YAN Chao-gui,FAN Miao,ZHOU Xu-hui,LI Shu-rong.Study of 64 slice detector carotid CT angiography with low tube voltage[J].Journal of Diagnostic Imaging & Interventional Radiology,2011,20(5):323-326.
Authors:ZHENG Shao-yan  CHU Jian-ping  YANG Zhi-yun  YAN Chao-gui  FAN Miao  ZHOU Xu-hui  LI Shu-rong
Institution:. Deportment of Radiology, The Center Hospital of Shantou City, 515031 Guangdong, China
Abstract:Objective To evaluate the clinical application of carotid CT angiography using a low tube voltage (100kV). Methods 55 patients requiring carotid CT angiography were randomly assigned to one of the two protocol groups: protocol A(27/55, scanning with 100kV and 150mAs); protocol B (28/55, scanning with 120kV and 150mAs). CT image quality including the signal intensity (SI), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and the radiation doses (CTDIvol and DLP) were compared between two groups. Kruskal-Wallis test was used for statistical analysis. Results The mean image noise value with protocol A (14.4 HU) was slightly higher (P 〈 0.0001) than that with protocol B (11.5 HU). The SNR with protocol A was slightly lower than that with protocol B. The mean SI and CNR revealed no significant difference between the two groups (SI: 303.6HU vs 301.7HU,P= 0.873; CNR: 17.8 vs 22.0,P= 0.142). Image qualitative analysis revealed no significant difference between the two groups. Compared with protocol B, the CTDIvol, DLP and ED of protocol A reduced by 33% (14 mGy vs 22 mGy), 38% (417mGy-cm vs 669 mGy-cm), 37% (2.25mSv vs 3.61 mSv) respectively. Conclusion Carotid CT angiography with 100kV in 64 Multi-detector CT can reduce radiation dose and get a good image quality.
Keywords:Computed tomography  Low dose CT  CT angiography  Multi-slice detector CT  Radiation dose