首页 | 本学科首页   官方微博 | 高级检索  
检索        


Study on the stability of the calibration and normalization in PET and the influence of drifts on the accuracy of quantification
Authors:Hans W Reist  Otto Stadelmann  Werner Kleeb
Institution:(1) Paul Scherrer Institute, CH-5234 Villigen, Switzerland
Abstract:With regard to the quality control of quantification in positron emission tomography some characteristics were examined, to develop a simple method for frequent monitoring. The stability and uniformity of the detector count rate was checked by plotting the RMS deviation of the non normalized count rate and the standard deviation of the normalized count rate, each normalized to its value after the calibration or normalization respectively. Switching off a single detector did not impair the image quality, but the normalized image pixel counts were reduced by 2%–3% when a detector block was switched off. Thus in case of need, a weak detector can still be used to perform a scan. A reduced count rate capability at specific activities above 5x104 Bq/ml (sim1.5x105 corrected true system counts/s or sim1.5x104/s for a plane) was found compared to the maximum usable activity of 8x104Bq/ml obtained 1 year earlier, indicating a drift in the count loss corrections. A variation of the room temperature changes the temperature distribution inside the gantry by 5% per °C and the drift of the sensitivity (normalized image pixel counts) is 2% per °C.This article was presented at the 1st EEC workshop on accuracy determination in PET, January 19–20th. 1989 Pisa, Italy (COMAC-BME Concerted Project ldquoCharacterization and Standardization of PET Instrumentationrdquo)
Keywords:Stability  Quality control  Temperature drift
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号