Validation of intensity modulated radiation therapy patient plans with portal images]
Authors:G Delpon  S Warren  D Mahé  S Gaudaire  A Lisbona
Institution:Service de physique médicale, CLCC Nantes Atlantique, Nantes Saint-Herblain, France. g-delpon@nantes.fnclcc.fr
Abstract:The goal of this study was to show the feasibility of step and shoot intensity-modulated radiation therapy pre-treatment quality control for patients using the electronic portal imaging device (iViewGT) fitted on a Sli+ linac (Elekta Oncology Systems, Crawley, UK) instead of radiographic films. Since the beginning of intensity-modulated radiation therapy treatments, the dosimetric quality control necessary before treating each new patient has been a time-consuming and therefore costly obligation. In order to fully develop this technique, it seems absolutely essential to reduce the cost of these controls, especially the linac time. Up to now, verification of the relative dosimetry field by field has been achieved by acquiring radiographic films in the isocenter plane and comparing them to the results of the XiO planning system (Computerized Medical Systems, Missouri, USA) using RIT113 v4.1 software (Radiological Imaging Technology, Colorado, USA). A qualitative and quantitative evaluation was realised for every field of every patient. A quick and simple procedure was put into place to be able to make the same verifications using portal images. This new technique is not a modification of the overall methodology of analysis. The results achieved by comparing the measurement with the electronic portal imaging device and the calculation with the treatment planning system were in line with those achieved with the films for all indicators we studied (isodoses, horizontal and vertical dose profiles and gamma index).
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号