首页 | 本学科首页   官方微博 | 高级检索  
     


A method for pneumatically inserting an array of penetrating electrodes into cortical tissue
Authors:Patrick J. Rousche  Richard A. Normann
Affiliation:(1) Department of Bioengineering, University of Utah, 84112 Salt Lake City, UT
Abstract:The goal of this research was to find a practical means by which an array of 100 needle-shaped electrodes could be implanted into the cerebral cortex with minimal brain tissue trauma. It was found that insertion of these structures into cortical tissues could only be performed using high insertion speeds. A pneumatically actuated impact insertion system has been developed that is capable of inserting an electrode array into feline brain tissue at speeds from about 1 to 11 m/s. We found that a minimum array insertion speed of 8.3 m/s was necessary for a complete, safe insertion of all 100 electrodes in the array to a depth of 1.5 mm into feline cortex. The performance of the impact insertion system is discussed in terms of a simplified representation of cortical tissue.
Keywords:Penetrating electrode array    Impact insertion    Cortical neurophysiology    Neuroprosthetics
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号