Single-acquisition chemical-shift imaging of a binary system with use of stimulated echoes |
| |
Authors: | Yeung, HN Kormos, DW Sebok, DA |
| |
Affiliation: | Technicare Corporation, Solon, Ohio. |
| |
Abstract: | A method for separating binary chemical-shift components with a single image data acquisition by means of stimulated echoes is demonstrated. With a strategy analogous to the modified Dixon method, three stimulated echoes were acquired to form three complex images. In each of the images, the complex pixel intensities were imparted, by design of the pulse sequence, with a phase factor carrying chemical-shift or field inhomogeneity information. With these three images, true fat/water separation can be obtained in biologic tissues. Studies at high field strength (4.7 T) on a toluene phantom, a pseudo-binary chemical-shift system, were used to evaluate the applicability of the method. Its clinical feasibility was demonstrated on a healthy human subject in a 0.6-T whole-body imaging system. |
| |