首页 | 本学科首页   官方微博 | 高级检索  
     


Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue
Authors:Lars Mattison PhD  Atul Verma MD  Khaldoun G. Tarakji MD  MPH  Tobias Reichlin MD  Gerhard Hindricks MD  Kevin L. Sack PhD  Birce Önal PhD  Megan M. Schmidt PhD  Damijan Miklavčič PhD  Daniel C. Sigg MD  PhD
Affiliation:1. Medtronic, Minneapolis, Minnesota, USA;2. McGill University Health Center, McGill University, Montreal, Quebec, Canada;3. Department of Cardiology, Inselspital—University Hospital Bern, University of Bern, Bern, Switzerland;4. Department of Electrophysiology, Heart Center Leipzig at University of Leipzig, Leipzig, Germany;5. Faculty of Electrical Engineering, University of Ljubljana, Ljubljana, Slovenia
Abstract:

Introduction

Contact force has been used to titrate lesion formation for radiofrequency ablation. Pulsed field ablation (PFA) is a field-based ablation technology for which limited evidence on the impact of contact force on lesion size is available.

Methods

Porcine hearts (n = 6) were perfused using a modified Langendorff set-up. A prototype focal PFA catheter attached to a force gauge was held perpendicular to the epicardium and lowered until contact was made. Contact force was recorded during each PFA delivery. Matured lesions were cross-sectioned, stained, and the lesion dimensions measured.

Results

A total of 82 lesions were evaluated with contact forces between 1.3 and 48.6 g. Mean lesion depth was 4.8 ± 0.9 mm (standard deviation), mean lesion width was 9.1 ± 1.3 mm, and mean lesion volume was 217.0 ± 96.6 mm3. Linear regression curves showed an increase of only 0.01 mm in depth (depth = 0.01 × contact force + 4.41, R2 = 0.05), 0.03 mm in width (width = 0.03 × contact force + 8.26, R2 = 0.13) for each additional gram of contact force, and 2.20 mm3 in volume (volume = 2.20 × contact force + 162, R2 = 0.10).

Conclusion

Increasing contact force using a bipolar, biphasic focal PFA system has minimal effects on acute lesion dimensions in an isolated porcine heart model and achieving tissue contact is more important than the force with which that contact is made.
Keywords:cardiac ablation  contact force  irreversible electroporation  pulsed field ablation